CD Skripsi
Kaji Eksperimental Blood Carrier Box Dengan Thermoelectric Cooler Kapasitas 2,45 Liter
ABSTRACT
The process of distributing blood from the proper storage before transfusion
requires a cooling unit that is able to maintain the temperature so that the blood
can be maintained and not damaged before being transfused. To maintain the
chemical and structural composition of blood, it is done by storing it at a
temperature of 2-6 °C and the distribution is carried at a temperature of 2-10 °C
with a maximum transit time of 24 hours specifically for whole blood (WB). To
answer this problem, this research will test the ability of a box that is used as a
container for the blood distribution process using a thermoelectric cooling system.
The thermoelectric used is type TEC1 12706 with the coefficient of performance
(COP) value based on the highest calculation result of 0.176. After testing, it
appears that the blood carrier box with the thermoelectric cooler running can
maintain the blood temperature, which is initially 2 oC until the end of the test for
2 hours ranging from 2-4 oC, then in testing the blood carrier box with the
thermoelectric cooler turned off can maintain the blood temperature. Initially 2 oC
until the end of the test for 2 hours ranged from 8-12 oC and in testing where the
blood carrier box with a thermoelectric cooler that was run could lower the blood
temperature from 24 oC to the end of the 2-hour test ranging from 16-22 oC.
Keywords: Blood Carrier Box, Thermoelectric Cooler, Coefficient of Peformance
Tidak tersedia versi lain